Boonton 4500B Peak Power Meter Bedienungsanleitung Seite 296

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 372
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 295
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-96
TRIGger:STATistical:GATE
Description: Set or return the marker locations in time relative to the trigger that define the time gate for gated
statistical measurements. These two gate markers function in the same manner as time markers in
the Pulse mode. See MARKer:POSItion[1|2]:TIMe <n> for more information.
Syntax: TRIGger:STATistical:GATE[1|2] <n>
Argument: <n> = display-time in seconds (see restrictions)
Valid Modes: Gated Statistical mode
Restrictions: For zero delay trigger position in the center of the display , the following relationship must be
satisfied: TrigDly - (5*time/div) < GateTime < TrigDly + (5*time/div) where the timebase
setting is time/div
TRIGger:STATistical:GATIng
Description: ing is needed.

accept only samples that fall within the gate defined by the Gate1 and Gate 2 markers.
Syntax: TRIGger:STATistical:GATIng <asc>
Argument: <asc> = {ON, OFF}
Valid Modes: Statistical mode