Boonton 4500B RF Peak Power Analyzer
Operation
Filter setting. This results in a "peak hold" characteristic where the min and max peaks increase immediately on the trace
when acquired and then, decay slowly. To hold the peaks longer, increase the Chan # > Exten > Filter setting. To maintain
the min and max peaks, turn Chan # > Exten > Peak Hold hannel.
In Statistical mode (Option 10 only), the CDF and CCDF are drawn filled in below the trace line when enveloping is on. The
PDF is always drawn as a filled distribution.
Note that the envelope setting does not affect any readings returned by markers, automeasure system, or remote commands.
Only the display presentation is affected.
Table 4-44. Display > Menu
Menu Item Function
(Type) Selections (SCPI cmd)
Persistence Pulse Enables or disables trace persistence.
(Toggle) Off, On See discussion above, Sect. 4-18.
DISPlay:PERSistence {ON, OFF}
Envelope Pulse, Mod & Stat Enables or disables trace envelope mode.
(Toggle) Off, On See discussion above, Sect 4-18.
DISPlay:ENVElope {ON, OFF}
Beep Off, On Controls the processor audible beeper
(Toggle)
is pressed.
SYSTem:BEEP {ON, OFF}
Brightness 10% to 100% Controls the display backlight
(Numeric)
Use any of the data entry keys to vary the
intensity of the LCD backlight. The default
setting is 90%.
DISPlay:BACKlight:BRIGhtness
<0 to 100>
Screen Saver MENU Opens the screen saver submenu
(Submenu)
Fig 4-60. Display >
Menu