Boonton 4500B Peak Power Meter Bedienungsanleitung Seite 127

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 372
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 126
Boonton 4500B RF Peak Power Analyzer
Operation
4-55
Table 4-12. Mark > Extensions > Submenu (continued)
Menu Item Function
(Type) Selections (SCPI cmd)
Each marker can be assigned to the memory channels also. All marker
functions will function on the memory channels including minimum
and maximum power measurements between markers and average
power between markers.
MARKer[1/2]:ASSIgn {ACTIVE, CH1, CH2, TRIG1, TRIG2,
CHM, MEM1, MEM2}
Delta Marker Pulse & Mod Selects the function of the center marker
(Mult. Choice) Ratio, Avg, window above the graphical display
Delta
Use Mark > Marker # menu function to place
the time markers in the active window at
the points of interest on the waveform. Press
the Mark > Delta Marker menu key to select

is available for non-logarithmic units only.)

voltage) ratio of Marker 1 and Marker 2 in
the center window. If Marker 1 and Marker 2
are expressed in logarithmic units, ratio will
be in dB. If Marker 1 and Marker 2 are
expressed in linear (non-logarithmic) units,
the ratio will be in percent.

(or voltage) in the waveform interval between
the two markers. The average power (or
voltage) will be expressed in the same units
as the Markers 1 and 2.

Mark > Extensions > Mk Group Both is
selected. When Mark > Extensions >
Mk Group Each is selected, the Delta Marker
selection is automatically switched to
Mark > Extensions > Mk
Group Both is reselected, the Delta Marker

Fig. 29b
The "Delta" selection is available only for Mark > Exten menu
linear (non-logarithmic) units (e.g. watts, Each CH1,Ch2
volts). The delta function displays the
power (or voltage) difference between
Seitenansicht 126
1 2 ... 122 123 124 125 126 127 128 129 130 131 132 ... 371 372

Kommentare zu diesen Handbüchern

Keine Kommentare