Boonton 4500B RF Peak Power Analyzer
Operation
4-117
AUTOMATIC MEASUREMENTS MODULATED MODE
The Modulated mode is most often used to measure the true average power of a signal on a continuous basis. The average
readings reflect the filtered average value. The peak readings are affected by the Chan # > Exten > Peak Hold setting. If peak
 pressing the Top Level > Display > CLEAR menu key. If peak
Chan # > Exten > Filter setting. For memory channels, the
peak value displayed is the peak value at the time the waveform was stored.
The peak-to-average reading is always in dB units and is not measured on External Trigger channels.
Table 4-59. Modulated Mode Automatic Measurements I
CH1&2 CH 1&2
Name (Power Units) (Voltage Units)
Avg Average power Average voltage (converted)
Peak Peak power Peak voltage
Min Minimum power Minimum voltage
Pk/Avg Peak to Average Ratio (dB) Peak to Average Ratio (dB)
Dyn Rng Dynamic range (dB) Dynamic range (dB)
Table 4-60. Modulated Mode Automatic Measurements II
EXT TRIG 1&2
Name (Voltage Units)
Avg Average voltage
Peak Peak positive voltage
Min Minimum voltage
Pk/Avg ----
Dyn Rng Peak-to-peak voltage