Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-64
5.5.13 REFLine Subsystem
The REFLine group of commands is used to configure and locate horizontal measurement lines (cursors) at specific levels on
the display. Reference lines can also be set to track waveform parameters and marker intersections.
REFLine:ASSIgn
Description: Set or return the assignment of Reference Line 1 or Reference Line 2 to a channel.
Syntax: REFLine:ASSIgn <asc>
Argument: <asc> = {OFF, ACTIVE, CH1, CH2, TRIG1, TRIG2, MEM1, MEM2}
Valid Modes: Any.
REFLine:POSItion:LEVel
Description: Set or return the vertical position (level) of Reference Line 1 or Reference Line 2 in channel
units..
Syntax: REFLine[1|2]:POSItion:LEVel <asc>
Argument: <asc> = 99.00 to +99.00 dBm (typical for dBm units)
Valid Modes: Any.
REFLine:POSItion:FORCetomk
Description: Force the position of Reference Line 1 and Reference Line 2 to the intersections, respectively, of
Marker 1 and Marker 2 
assigned to a different channel or does not intersect the waveform, the corresponding reference
 command.
Syntax: REFLine:POSItion:FORCetomk
Argument: None
Valid Modes: Any.