Boonton 4500B RF Peak Power Analyzer
Operation
4-68
4.15 UTIL Key and Util > Menu
The UTIL key activates the Util > menu (See Table 4-20 and Figure 4-26), which enables you to view the instrument status
summary, set up the IEEE-488 bus and serial input/output ports, and set the internal real-time clock.
Table 4-18. Util > Menu
Menu Item
(Type) Selections Function
Inst Status REPORT Displays equipment serial numbers,
configuration and revision status.
Press the Util > Inst Status menu key to
display the serial number, configuration,
and revision status. See Figure 4-34. If
Channel 1 or 2 is not responding, a channel
card is either not functional or not installed.
If the Instrument Status Report indicates that
the sensor is not connected, check the cable
and sensor connections. The Report will
also indicate the cable length for the input
board. The cable length should match the
configuration for the input board.
Remote Conf MENU Opens the Util > Remote Conf submenu
Press the Util > Remote Conf menu key to
open the remote configuration submenu.
Disk MENU Opens the Util > Disk submenu
Hardcopy MENU Opens the Util > Hardcopy submenu
Clock MENU Opens the Util > Clock submenu
Fig.4-32. Util > menu
Fig. 4-34. Util > menu