Time base: Boonton software uses a fixed grid of 10 divisions
for the time-scale. The time resolution is set per division with
the lowest possible value of 5 ns/div.
Trigger Delay: Trigger delay can be adjusted both manually
entering numerical value into the field or using up-down
arrow keys.
Trigger Position: Trigger position can be changed by entering
numerical values into the field and also by dragging a
horizontal bar toward three default positions: Left, Middle
and Right.
Trigger sources and modes: Several trigger modes available
for each trigger source. Multiple trigger sources available
under the drop down list including both "Internal" and
"External" selection.
Trigger Mode: Four trigger modes available: Normal, Auto,
AutoLevel and Free run.
Trigger Level: Trigger level can be changed by using arrow
keys
Slope: Trigger slope can be selected by using positive and
negative options under slope.
Holdoff: Holdoff position allows user to select two different
modes: Normal and Gap.