INSTRUCTION MANUAL 4240 SERIES RF POWER METER REV DATE 01/19/2011 MANUAL P/N 98406700A 06799A CD P/N 984

Seite 2

Boonton 4240 Series RF Power Meter This page intentionally left blank. Contents x

Seite 3 - SAFETY SUMMARY

Boonton 4240 Series RF Power Meter 5-1. SCPI COMMAND SUMMARY (continued) SENSe[1|2]:CORRection:CA<numeric value> = -3.00 to 3.00 et/return dut

Seite 4

Boonton 4240 Series RF Power Meter Table 5-1. SCPI COMMAND SUMMARY (continued) STATus:QUEStionable:CALibration:EVENt? Ret urn value

Seite 5

Boonton 4240 Series RF Power Meter 5.5.16 4230 Emulation GPIB Commands Table 5-2. 4230 EMULATION GPIB COMMANDS Code Description Comments CP

Seite 6

Boonton 4240 Series RF Power Meter Table 5-2. 4230 EMULATION GPIB COMMANDS (continued) Code Description Comments Channels ngle channel al, fr

Seite 7

Boonton 4240 Series RF Power Meter 5.5.17 HP 437B Em ulation GPIB Commands Table 5-3. HP 437B EMULATION GPIB COMMANDS UPPORTED RG # 100) NOT

Seite 8

Boonton 4240 Series RF Power Meter Table 5-3. HP 437B EMULATION GPIB COMMANDS Code D s KZ ction function NOT SUPPORTED llator off n t

Seite 9

Boonton 4240 Series RF Power Meter Table 5-3. HP 437B EMULATION GPIB COMMANDS Code Description Comments SM Status message NOT SUPPOR

Seite 10 - Contents

Boonton 4240 Series RF Power Meter 5.5.1 ands 8 HP 438A Emulation GPIB CommTable 5-4. HP 438A EMULATION GPIB COMMANDS Comments D Measure A-

Seite 11 - 1. General Information

Boonton 4240 Series RF Power Meter LP1 Set learn mode #1 NOT SUPPORTED Table 5-4. HP 438A EMULATION GPIB COMMANDS ode Description Comm

Seite 12

Boonton 4240 Series RF Power Meter 6. Application Notes in this section include pulse measurement fundamentals, automatic measurement rinciples, and a

Seite 13 - General Information

Boonton 4240 Series RF Power Meter 1. General Information This instruction manual provides you with the information you need to install, operate and

Seite 14

Boonton 4240 Series RF Power Meter Figure 6-1. Pulsed RF Signal Figure 6-2. Distorted Pulse Signal Application Note

Seite 15

Boonton 4240 Series RF Power Meter 6.1.2 Diode Detection Wideband diode detectors are the dominant power sensing device used to measure pulsed RF

Seite 16 - Sensor dependant )

Boonton 4240 Series RF Power Meter Frequenccapacicannottable f Misma nding on the mismatminim Signalcalibrat nal and second harmonic combininsigni egi

Seite 17

Boonton 4240 Series RF Power Meter 6.2 M The 424 Standard year calibratio chapter assumes that the power meter is being maintained correctly and is

Seite 18

Boonton 4240 Series RF Power Meter Tual his reason, the uncertainties are more commonly combined using the RSS method. RSS is an abbreviation for “ro

Seite 19

Boonton 4240 Series RF Power Meter The value to use for calibration level uncertainty depends upon the sensor calibration technique used. If AutoCal

Seite 20

Boonton 4240 Series RF Power Meter Sensor Temperature Coefficient. This term is the error which occurs when the sensor’s temperature has changed sign

Seite 21 - 2. Installation

Boonton 4240 Series RF Power Meter 6.2.3 Sample Uncertainty Calculations. The following example shows calculations for a CW power sensors. The fig

Seite 22 - Installation

Boonton 4240 Series RF Power Meter Step 6: The Sensor Temperature Drift Error depends on how far the temperature has drifted from the sensor calibrati

Seite 23 - 2.4 Preliminary Check

Boonton 4240 Series RF Power Meter Step 1 certainty terms has been determined, we can combine them to calculate the worst-case a U (±%) K (U×K)2

Seite 24

Boonton 4240 Series RF Power Meter 1.2 Description The Model 4240 is a digital signal processor based, single or dual channel, solid state RF power

Seite 25 - 3. Getting Started

Boonton 4240 Series RF Power Meter This page intentionally left blank. Application Notes 6-12

Seite 26 - Getting Started

Boonton 4240 Series RF Power Meter 7. Maintenance Th .1 Safety

Seite 27

Boonton 4240 Series RF Power Meter 7.5 Firmware Upgrade Instructions equirements The Firmware Update file is a text file in S-record format. Any uti

Seite 28

Boonton 4240 Series RF Power Meter 7. During time the 4240 will display "FIRMWARE DOWNLOAD". Thapprox e 45 seconds to completely transfer.

Seite 29

Boonton 4240 Series RF Power Meter This page intentionally left blank. Maintenance 7-4

Seite 30 - 3.3 Operation

Boonton 4240 Series RF Power Meter 8. ges Appendix A SCPI Error Messa .1 SCPI Erro8 r Messages -101 “SubCmd not found” -113 "Undefined head

Seite 31

Boonton 4240 Series RF Power Meter NO. MESSAGE DESCRIPTION sor Disconnected." -249 "FPGA download err" 63 “MFS Init” 64 “Flash i

Seite 32

Boonton 4240 Series RF Power Meter Appendix B 9-1 9. Appendix B Warranty & Repair Repair Policy 4240 Series Instrument. If the Boonton 4240 Ser

Seite 33

Boonton 4240 Series RF Power Meter Appendix B 9-2 THE FOREGOING WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUD-ING, BUT

Seite 34

Boonton 4240 Series RF Power Meter • ual Independent Channels. When equipped with the optional second measurement channel, the • • optimized throu

Seite 35

Boonton 4240 Series RF Power Meter • • ne key press operations. To provide for ease of use operation of the instrument functions that mply by pres

Seite 36

Boonton 4240 Series RF Power Meter 1.4 Accessories 40 ac data sensor adapter for each channel installed the instrument. One or more Boonton 51000

Seite 37

Boonton 4240 Series RF Power Meter Option designations are appended to the instrument’s base model number. For example, Model 4242-01-nel instrum ca

Seite 38

Boonton 4240 Series RF Power Meter Table 1-2 4240 Series Performance Specifications (continued) pecifications aNCERTAINTIES easurement Accuracy Sum

Seite 39

Boonton 4240 Series RF Power Meter Table 1-2 4240 Series Performance Specifications (continued) (Specifications are subject to change without notic

Seite 40

Boonton 4240 Series RF Power Meter Table 1-2 4240 Series Performance Specifications (continued) (Specifications are subject to change without notic

Seite 41

Boonton 4240 Series RF Power Meter INSTRUCTION MANUAL, 4240 SERIES RF POWER METER Revision date 01/19/2011 © Cop

Seite 42

Boonton 4240 Series RF Power Meter This page intentionally left blank. General Information 1-10

Seite 43

Boonton 4240 Series RF Power Meter 2. Installation This section contains unpacking and repacking instructions, power requirements, connection descrip

Seite 44

Boonton 4240 Series RF Power Meter Table 2-1 4240 Series Packing List INSTRUMENT (See also Table 1-1) 4240 Series RF Power Meter Line Cord Boon

Seite 45

Boonton 4240 Series RF Power Meter Recorder If a recorder is to be used to record measurement data, connect the recorder to the recorder BEC conne

Seite 46

Boonton 4240 Series RF Power Meter 6. Press the <MENU> key and select DIAGNOSTICS with the down arrow key. Press <ENTER>.

Seite 47 - 4. Operation

Boonton 4240 Series RF Power Meter 3. Getting Started This chapter will introduce the user to the 4240 Series. The chapter will identify objects on

Seite 48 - Operation

Boonton 4240 Series RF Power Meter Figure 3-1. Standard 4240 Series RF Power Meter - Front Panel Table 3-1 Operating Controls, Indicators and C

Seite 49

Boonton 4240 Series RF Power Meter Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear Nomenclature Fu

Seite 50

Boonton 4240 Series RF Power Meter Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear Nomenclature Fu

Seite 51 - 4.5 Noise

Boonton 4240 Series RF Power Meter Figure 3-2. 4240 Series - Rear Panel (Shown without optional rear panel connectors not installed) Getting Star

Seite 52

Boonton 4240 Series RF Power Meter SAFETY SUMMARY The following general safety precautions must be observed during all phases of operation and main

Seite 53

Boonton 4240 Series RF Power Meter 3.3 Operation The Model 4 c e r e o i e c h o l 240 an b configu ed for op rati n v a th six swit hes on

Seite 54

Boonton 4240 Series RF Power Meter DUAL CHANNEL ACM M M M M L U PK∆ ± D D D D D D U U U ( B A R G R A P H )

Seite 55

Boonton 4240 Series RF Power Meter 3-5. Mod nd Set ast Menu Operatio ey stokes to perform a function repeat d , the control program can ing to t .

Seite 56 - .6 Measurement Time

Boonton 4240 Series RF Power Meter 3.3.1 Menu Key. he instrument's, CHA P, REPORT and DIAGNOSTIC functions are accessed when the sor can be posit

Seite 57 - 4.7 High Frequ

Boonton 4240 Series RF Power Meter Table 3-2 CHANNEL MENU Functions Function De Parameters Defaults RTN e n e tr e o r u e . NITS Units used for

Seite 58 - 4.8 Waveform Sensitivity

Boonton 4240 Series RF Power Meter R T N L E L 6 V E > - 0

Seite 59

Boonton 4240 Series RF Power Meter Table 3-4 SETUP MENU Functions n u Function De rs Defaults SAVE Saves the current instrument 1- 0 1 confi

Seite 60 - 4.10 Bar Graph Operation

Boonton 4240 Series RF Power Meter Table 3-5 IEEE MENU Functions (continued) OS LSTN End of string indicator for LF, CR, CRLF, NONE LF received m

Seite 61 - 5. Remote Operation

Boonton 4240 Series RF Power Meter Table 3-6 RS232 MENU Functions (continued) Function Description rs e u PARITY bit mode in a m ssage. ODD, NO

Seite 62 - 5.3 SCPI Language

Boonton 4240 Series RF Power Meter Table 3-7 DIAGNOSTICS MENU Functions (continued) Function Description Parameters Defaults SELF TEST Instructs

Seite 63

Boonton 4240 Series RF Power Meter SAFETY SYMBOLS This safety requirement symbol (located on the rear panel) has been adopted by the International El

Seite 64

Boonton 4240 Series RF Power Meter Figure 3-15. Access Code R T N M &

Seite 65

Boonton 4240 Series RF Power Meter Function Description Parameters Defaults : Factor 5000 [0 to 6] : [0 to 9999] the n/a n/a calibratio

Seite 66 - 5.5 SCPI Command Reference

Boonton 4240 Series RF Power Meter Table 3-9 FREQ Menu Functions unction Description Parameters Defaults F FREQ Frequency 0.01 to 100.00 GHz

Seite 67

Boonton 4240 Series RF Power Meter 3.3.3 FREQ Key. he freTinquency of the signal being measured must be entered in order to use the stored high frequ

Seite 68

Boonton 4240 Series RF Power Meter he user may also perform a 0 dBm calibration by one key stT roke of the Zero/Cal key. Simply connect the sensor to

Seite 69

Boonton 4240 Series RF Power Meter Table 3-10 Reference Level Menu Functions unction Description Parameters Defaults Bm Reference level value in

Seite 70

Boonton 4240 Series RF Power Meter This page intentionally left blank. Getting Started 3-22

Seite 71

Boonton 4240 Series RF Power Meter 4. Operation This section provides detailed background information on various aspects of operation of the Model 4

Seite 72

Boonton 4240 Series RF Power Meter these figures show the decay of measured power after a large signal has been applied. The typical error that can b

Seite 73

Boonton 4240 Series RF Power Meter Figure 4-2 Diode Sensor Decay Operation 4-3

Seite 74

Boonton 4240 Series RF Power Meter 1. General Information...

Seite 75

Boonton 4240 Series RF Power Meter 4.3 Dynamic Range The hold range mode is useful when he hold range mode is active when a specific instrument e

Seite 76

Boonton 4240 Series RF Power Meter 4.4 Filtering The Model 4240 employs digital filtering (averaging of measurements) to reduce the noise floor of

Seite 77

Boonton 4240 Series RF Power Meter Operation 4-6

Seite 78

Boonton 4240 Series RF Power Meter Figure 4-5. Noise Reduction Figure 4-6. Typical Error Band Specifications Figure 4-7. Probability of Falling wi

Seite 79

Boonton 4240 Series RF Power Meter Figure 4-8. Confidence Curves, 51013 Sensor with 2.8 Second Filter Operation 4-8

Seite 80

Boonton 4240 Series RF Power Meter Figure 4-9. Confidence Curves, 51013 Sensor with 10 Second Filter Operation 4-9

Seite 81

Boonton 4240 Series RF Power Meter Noise Error Examples. Figures 4-8 and 4-9 show the computed error for the 51013 diode sensor at different power le

Seite 82

Boonton 4240 Series RF Power Meter ontinuous Response. Regardless of the overhead time or the digital filter length, the Model 4240 will output readi

Seite 83

Boonton 4240 Series RF Power Meter Figure 4-11. Mismatch Uncertainties Chart 4.8 Waveform Sensitivity Thermal sensors are insensitive to the wav

Seite 84

Boonton 4240 Series RF Power Meter Diode 20 dBm (-10 dBm and 0 dBm for attenuated models 51015 and 1033). This response characteristic is obtained b

Seite 85

Boonton 4240 Series RF Power Meter 4.3 Dynamic Range...

Seite 86

Boonton 4240 Series RF Power Meter nly the display is affected by the duty cycle calculation. The mOperation 4-14 easurement process is subject to t

Seite 87

Boonton 4240 Series RF Power Meter 5. Remote Operation 5.1 GPIB Configuration The 4240 GPIB interface is configured using the Menu > SETUP >

Seite 88

Boonton 4240 Series RF Power Meter Additionally, the ASCII "DC2" character (hexadecimal 12/CTRL R) will cause the instrument to immediately

Seite 89

Boonton 4240 Series RF Power Meter suffix is not present, the default channel is assumed. strument’s “Channel 1” measurement path, while SENSFor exam

Seite 90

Boonton 4240 Series RF Power Meter Examples: :CALCulate:STATe ON Turn on measurement channel 1 (default channel number) ON Turn on measurement chan

Seite 91

Boonton 4240 Series RF Power Meter 5.4 Basic Measurement Information e easiest asequence which includes a MEAS1:POWERMEAS1:VOLTAGE? To return th

Seite 92

Boonton 4240 Series RF Power Meter 5.5 SCPI Command Reference te commands accepted by the 4240 Series. The list is grouped by SCPI subsystem r IEEE

Seite 93

Boonton 4240 Series RF Power Meter *ESR? Description: Return the current value of the Standard Event Status Register, then clear the register. Th

Seite 94

Boonton 4240 Series RF Power Meter *OPT? Description: Return the status of Channel 1 and Channel 2 followed by a list of installed options. Syntax

Seite 95

Boonton 4240 Series RF Power Meter *STB? Description: Return the current value of the Status Byte register. This register has bits assigned to a

Seite 96

Boonton 4240 Series RF Power Meter CALCulate:LIMit:LOWer[:POWer] ... 5-10 CALCulate:LIMi

Seite 97

Boonton 4240 Series RF Power Meter 5.5.2 CALCulate Subsystem he CALCulate group of the command subsystem is used to configure post acquisition data p

Seite 98

Boonton 4240 Series RF Power Meter CALCulate:LIMit:UPPer[:POWer] Description: Set or return the upper limit power level for the selected channel. Thi

Seite 99

Boonton 4240 Series RF Power Meter CALCulate:MATH:ARGA Description: Set or return the first argument to be used for channel math operations.

Seite 100 - Remote Operation

Boonton 4240 Series RF Power Meter CALCulate:REFerence:COLLect Description: For the selected channel, make the current power the reference level

Seite 101

Boonton 4240 Series RF Power Meter 5.5.3 CALibration Subsystem ommands is used to control automatic zero offset and linearity adjustments to the RF p

Seite 102

Boonton 4240 Series RF Power Meter 5.5.4 DISPlay Subsystem Syntax: DISPlay:ACTive <numeric value> Argument: <numeric value> = 1 to 2

Seite 103

Boonton 4240 Series RF Power Meter DISPlay:LABel:TEXTA Description: Displays a text message of up to 20 characters in the first label field when Di

Seite 104

Boonton 4240 Series RF Power Meter 5.5.5 FETCh Queries The FETCh? group of queries is used to return specific measurement data from a measurement cyc

Seite 105

Boonton 4240 Series RF Power Meter 5.5.6 INITiate and ABORt Commands tion in progress and prepares e instrument for an INITiate command. BORt on:

Seite 106

Boonton 4240 Series RF Power Meter 5.5.7 MEASure Queries used to acquire data using a set of high level instructions. They are structured to allow

Seite 107

Boonton 4240 Series RF Power Meter OUTPut:LEVel[:POWer] ... 5-21 OUTPut

Seite 108

Boonton 4240 Series RF Power Meter 5.5.8 MEMory Subsystem is used to save and recall instrument operating configurations, and to edit and review use

Seite 109 - 6. Application Notes

Boonton 4240 Series RF Power Meter 5.5.9 OUTPut Subsystem The OUTPut group of commands is used to control various outputs of the 4240. These output

Seite 110 - Application Notes

Boonton 4240 Series RF Power Meter OUTPut:RECorder:SOURce Description: Set or return the source channel for the Recorder Output. Syntax: OUTPut:R

Seite 111

Boonton 4240 Series RF Power Meter 5.5.10 READ Queries he purp AData. READ s theis equivalent to ABORtinready. READ? generally every time it is iss

Seite 112

Boonton 4240 Series RF Power Meter 5.5.11 SENSe Subsystem The purpose of the SENSe command subsystem is to directly configure device specific setting

Seite 113

Boonton 4240 Series RF Power Meter SENSe:FILTer:STATe Description: Set or return the current setting of the integration filter on the selected cha

Seite 114

Boonton 4240 Series RF Power Meter 5.5.12 STATus Commands ables you to control the SCPI defined status reporting structures. The user may amine the

Seite 115

Boonton 4240 Series RF Power Meter STATus:DEVice:EVENt? Description: Returns the current contents of the Device Event register then resets the reg

Seite 116

Boonton 4240 Series RF Power Meter STATus :CO Description: real time, and can change es in the instrument’s operating condition. :OPERation NDition

Seite 117

Boonton 4240 Series RF Power Meter STATus:OPERation:NTRansition alue of the negative transition filter bitmask for the Operation status group. the n

Seite 118

Boonton 4240 Series RF Power Meter SYSTem:COMMunicate:SERial:PARity... 5-34 SYSTem:COMMunicate:

Seite 119

Boonton 4240 Series RF Power Meter STATus:QUEStionable:CONDition? f the Questionable Condition register. The following table shows the Bit Va Descr

Seite 120

Boonton 4240 Series RF Power Meter STATus:QUEStionable:NTRansition Description: Set or return the value of the negative transition filter bit mask

Seite 121 - 7. Maintenance

Boonton 4240 Series RF Power Meter STATus:QUEStionable:CALibration:ENABle Description: Sets or returns the Questionable Calibration Enable register

Seite 122 - Maintenance

Boonton 4240 Series RF Power Meter 5.5.13 SYSTem Subsystem T Tem group o mmands is used to control system-level functions not directly related to ins

Seite 123

Boonton 4240 Series RF Power Meter SYSTem:COMMunicate:GPIB:TALK of-string terminator . UD UD <numeric value> 00, 1200, 2400, 4800, 9600, 19200,

Seite 124

Boonton 4240 Series RF Power Meter SYSTem:ERRor[:NEXT]? d error code number followed by a quoted ASCII text string describing the error. Note that er

Seite 125 - Appendix A SCPI Error Messa

Boonton 4240 Series RF Power Meter .5.14 INSTrument:VERSion Commands p of commands is used to q ode follows a “YYYYMMDD” format. The FPGA revisio s

Seite 126 - Appendix A

Boonton 4240 Series RF Power Meter .5.15 SCPI Command Summary Table 5-1. SCPI 5 COMMAND SUMMARY ESE Set/get Standard EvenBORt to idle iate] L?

Seite 127 - HTwww.boonton.comTH

Boonton 4240 Series RF Power Meter Ta d) ble 5-1. SCPI COMMAND SUMMARY (continue CALCulate:MATH:ARGB Set or return the second argument to be used fo

Seite 128 - Appendix B

Boonton 4240 Series RF Power Meter Table 5-1. SCPI COMMAND SUMMARY (continued) DISPlay:Label:TEXTC Displays a text message in the third label field.